Shimadzu Dry-type Helium Leak Detector
This compact and transportable, general-purpose helium leak detector comes with a built-in contactless dry pump suitable for use in a clean room, it can be used for a variety of applications ranging from equipment maintenance to permanent installation such as OEM integration.
■Specifications MSE-2401
Sensitivity | 10-13~100Pa・m3/s[He] |
---|---|
Detecton gas | Hydrogen/Helium gas |
Analyzer tube | 270̊ magnetic deflecting type (Flux density : 0.17T) Yttria-coated iridium filament (2 pcs.) Automatic tuning sensitivity calibration with temperature compensation |
Vacuum System | Molecular drag pump specially designed for leak detector (Structure of exhaust 70L/s) Non-contact duty dry vacuum pump specially designed for leak detector (110L/min) Pirani gage : Provided in two lines |
Test Port | Maximum pressure 1000Pa Connection port : NW25KF Helium Pumping Speed 3L/s |
Weight | Main unit : approx. 59kg |
Power Requirements | Specify one of the following when placing an order : Single phase 100/110/115/120/200/220/230/240VAC ±10%, 50/60Hz, 700VA |
Input/output | RS-232C interface EXT-UNIT interface (external unit) |
■Dimensions MSE-2401
Application Example |
| Application Example | |
Vacuum Spray Method (work of approx. 5 litter)
Hood Method
Sniffer Method (Optional sniffer unit is necessary)
Bombing Method
(test chamber and bombing machine are necessary)
※ Test chamber and bombing equipment are required.
Chamber Method
Parallel Evacuation Method (The leak indicator unit is an option)
※ Wireless leak indicator is optional.